Breast Biopsy System Description:


In the MedWOW’s catalog, Breast Biopsy System is described as follows:
Biopsy units, some mammographic, used to take samples from breast tissue for testing.
The above ads list companies that sell and / or service Breast Biopsy System.

Breast Biopsy System may also be referred to as:


Biopsy Kits, Mammography | Breast Lesion Localization System | Mammographic Biopsy Stereotactic System | Mammography Unit | Stereotactic Biopsy, Mammography | Stereotactic Mammography Biopsy

Breast Biopsy System can be found under the following headings:


Oncology | Imaging | Surgery

Tips for buying Breast Biopsy System


  • Only stereotactic mammography biopsy systems that are covered by a service contract will get extended system performance and uptime guarantees beyond the length of the warranty from some suppliers.
  • Suppliers offer service contracts or service on a time-and-materials basis. This may also be available from a third-party organization. Facilities should carefully consider whether to purchase such a service, which will ensure the performance of preventive maintenance at regular intervals, minimizing unexpected maintenance costs.
  • If facilities need multiple stereotactic mammography biopsy systems or stereotactic breast biopsy systems to handle the patient volume, they should consider buying all units from one supplier for cost savings. The standardization of equipment within a facility simplifies staff training, servicing and parts acquisition, and provides greater bargaining leverage when negotiating costs.
  • Another important factor is modulation transfer function. MTF refers to the loss of contrast relative to an x-rayed object. As spatial frequencies in a given image increase, MTF decreases, creating loss of visualization. a 50% MTF at 5 line pairs per millimeter is recommended.
  • In digital detector performance, detective quantum efficiency is an important factor. DQE is directly related to the signal-to-noise ratio that results between detector input and output. The noise generated by the detector and the spatial resolution contributes to the DQE of the system. To reduce DQE, SNR must be reduced; a recommended DQE is less than 20% at 5 lp/mm.
  • Add-on stereotactic mammography biopsy systems may be harder to use, because they have more limited motions compared to dedicated stereotactic breast biopsy systems. Consequently, most of the differences are a matter of user preference, rather than clinical efficacy.
  • Film-based units are not recommended because they do not allow real-time guidance. Instead, both dedicated and add-on mammographic biopsy stereotactic systems are equipped with small digital detectors.
  • Facilities considering the purchase of a stereotactic breast biopsy system should consider the number of procedures to be performed and the compatibility with existing equipment.
  • To clearly image the smallest micro calcifications, SID of at least 66 cm is needed.
  • Focal spot sizes should be approximately 0.1-0.3 mm to ensure that micro calcifications as small as 200 µm can be detected.
  • AEC provides the appropriate image optical density and x-ray exposure for breast composition and thickness.
  • A rotating anode, which has a higher heat capacity than a stationary anode, should be available in the stereotactic breast biopsy system to promote longer x-ray tube life and provide a more consistent x-ray output.
  • The stereotactic mammography biopsy system should offer at least 500 mAs and an exposure time of 0.1 second or shorter to avoid unnecessary long exposures.
  • Mammography involves a narrow range of tissue thickness, and therefore 1 kV increments and a range of approximately 22 to 35 kV are necessary in a mammography unit.
  • A high-frequency x-ray generator will help ensure higher efficiency of operation with a minimum of output ripple. It will also need less space than a conventional generator.
  • When purchasing a mammography unit, performance and reliability are important factors.
  • Some manufacturers have developed advanced grids, which are expensive to manufacture but improve the image quality.
  • The complexity of the automatic exposure control is the main difference between systems. Some systems control only the exposure time, while more advanced ones control the x-ray spectrum, including the kVp, anode, and filtration. Such stereotactic mammography biopsy systems improve the image quality, especially in larger and denser breasts.
  • Facilities should base their mammographic biopsy stereotactic system purchase decision on life-cycle cost, local service support, discount rates and non-price-related benefits offered by the supplier, and standardization with existing equipment.
  • Stereotactic Mammography Biopsy screening equipment must meet the minimum MQSA requirements for tube outputs, compression performance, position controls, and reproducibility.